For more information about our products, please contact us
The ActivResonance Controller is available in two versions:
INF-102(A)
1-400khz
For operation in air only
INF-102(AL)
1-400khz
For operation in air (1- 400kHz) and liquid (1-50kHz)
Minimum reference input:
2 mV rms (equating to an actual drive of 0.66 mV rms)
Phase resolution:
0.1°
Phase variation:
maximum 5.0° / 10kHz
User Interface:
Phase: Digital Gain: Analogue
Conformity:
CE compliant
Dimensions:
345mm wide x 140mm high x 345mm deep
Weight:
7.75 kg
Mains Supply Voltage
IEC
230 V ac, 50 Hz, 40 VA
or 115 V ac, 60 Hz, 40 VA
Input
37-way D type DIN
Output
37-way D type DIN
Sensor input
BNC
0 V ptp to 10 V ptp, 0 V offset,
1 kHz to 400 kHz
Control input
BNC
3 mV ptp to 10 V ptp, 0 V offset,
1 kHz to 400 kHz
Output
BNC
0 V ptp to 10 V ptp, 0 V offset,
1 kHz to 400 kHz
2 additional outputs for monitoring various signals in the feedback circuit
BNC
0 V peak-peak to 10 V peak-peak
0v offset, 1 khz to 400 khz
The ActivResonance Controller is fitted with 37-way D and general purpose BNC connectors. This gives the flexibilty to interface the ActivResonance Controller to any scanning probe microscope, whether the cantilever is being driven acoustically or magnetically.
A simple software upgrade allows the user to upgrade an ActivResonance Controller A to an AL when required.